miniXS Calibration
------------------

In order to calibrate the miniXS spectrometer, an energy value needs to be assigned to each pixel of the Pilatus camera. To do this, several exposures should be taken with the incident beam set to various energies within the range of the spectrometer. The elastically scattered beam will then show up on the camera as a thin arc for each crystal. From the respective energies of these arcs, we can fit a smooth function to assign energy values to all pixels on the camera face.

Ideally, the calibration data should be taken by setting up a 1-d scan that triggers the Pilatus. Lets assume we have a scan file named "calibration.0001" with 15 energy points evenly spaced over the range of the spectrometer. The corresponding are saved in files calibration_00001.tif through calibration_00015.tif.

Starting
--------
Load the calibrator user interface.

Choose a name for the calibration and enter it in the Dataset Name box.

Selecting input files
--------------------
Click "Read Energies..." and in the dialog that pops up click on the "..." button and select the calibration.0001 scan file.

The "Energy Column" dropdown is automatically filled in with the names of the columns. The first column containing 'energy' (case insensitive) in its name is automatically selected for you. If the incident beam energy values are in a different column, select it from the dropdown.

When you click "Ok", the energy values are added to the first column of the list.

Now click on "Select Exposures". Go to the directory containing the .tif files and click once on the first one (calibration_00001.tif). Scroll down so that the final one is visible (calibration_00015.tif). Hold down shift and click on it. The entire set should now be selected. Click "Open", and the filenames will be added to the second column of the list.

If needed, you can add more energies or exposures to the end of the list by repeating the above steps.

Viewing the calibration exposures
---------------------------------

As soon as at least one row in the list contains an energy and an exposure file name, the first exposure will be loaded an displayed below the list.

To the left of the exposure view is a set of filters than can be applied.

# explain filters

Slide the slider under the view to switch between the loaded exposures.

Selecting Crystal Boundaries
----------------------------

Before calibration can proceed, we need to define the boundaries of spectra from each individual crystal in the spectrometer.

To do so, click in the exposure view and drag to draw a rectangle. Once drawn, the rectangle can be moved by clicking inside it and dragging, or resized by clicking near one of its borders and dragging.