# digiBase
Python interface to ORTEC/AMETEK digital MCA PMT base.

This is a 100% Python rewrite of C library interfaces:

* [libdbaserh](https://github.com/kjbilton/libdbaserh)
* [libdbase](https://github.com/SkyToGround/libdbase)

If you don't want to use the AMETEK Connections library, or can't, you may find this useful. 
I wrote this interface because I needed to create a Raspberry Pi data acquisition system for
use in remote locations. The functionality is basic but currently supports:

* HV programming (no readback yet)
* ADC _fine_ gain setting
* Livetime / realtime setting and reading
* Lower-level discriminator set
* EXT gate OFF, ENABLED, COINCIDENCE modes
* PHA mode
* List-mode acquisition

This connects to ORTEC/AMETEK digiBases over the USB bus and supports devices with USB vendor ID = 0x0a2d and product ID = 0x000f or 0x001f, which present slightly different communication
interfaces.

## Dependencies

* pyusb
* NumPy (I will try to refactor this out)

## Installation
Currently the code exists as a single python module. If using as a module, set PYTHONPATH as appropriate. The module also sports a command line interface that is fairly well documented using Python's argparse package:

```bash
$ ./digibase.py --help
```

will provide help on the various commands and options it supports.

### DigiBase Firmware
Device firmware must be loaded at power up. ORTEC distributes this firmware with
their MAESTRO and Connection software. You may find this in the distribution media
or in the install directories of the aforementioned software, in `\WINDOWS`, or 
in `\WINDOWS\SYSTEM32` with filename `digiBase.rbf` or `digiBaseRH.rbf`. Place this
file in the current working directory or set the environment variable 
`DBASE_FIRMWARE` to point to the file.

### USB Device Permission (Linux)
Under Linux the module may fail to open the USB device due to lack of permission. 
While it's always possible to `sudo chown 666 /dev/bus/usb/xxx/yyy` it will likely
be reset across power off or even after unplugging and plugging the device back in.
The better solution in this case is to add a custom udev rule:

```bash
sudo echo <<EOF > /etc/udev/rules.d/50-usb-perms.rules
SUBSYSTEM=="usb", ATTRS{idVendor}=="0a2d", GROUP="users", MODE="0666"
EOF
```

## Basic Usage - Python Module

This currently (I think) only supports one device so if you have multiple devices connected I think it will find the first one. ORTEC seems to have manufactured several versions 

```python
from digibase import digiBase, ExtGateMode
from time import sleep
base = digiBase()
print('Opened digiBase, serial number', base.serial)
```

should open the device and provide a serial number. 

### PHA Mode Acquisition
To setup a run in PHA mode you would need to configure settings:

```python
base.hv = 800
base.enable_hv()
base.lld = 24
base.ext_gate(ExtGateMode.COINCIDENCE)
base.realtime_preset = 15.0 # to set for 15 seconds
base.livetime_preset = 15.0 
base.set_acq_mode_pha()

sleep(5)        # Sleep 5 seconds to allow HV to stabilize 
base.start()    # Start the acquisition
```

The run should automatically stop after 15 sec since you programmed a preset above.
Note that the device does not block so you can stop the acquisition early if desired.

```python
sleep(5)
base.stop()
```

In any case you need to turn off the acquistion before starting again.

To access the pulseheight spectrum / MCA channels:

```python
spectrum = base.spectrum
```

which returns a python `array.array` type. It will be 1024, 32-bit unsigned integers.

### List Mode Acquisition
I find the so-called _list mode_ acquisition quite powerful. Instead of having logic on
the base fill histogram bins with the ADC values you get the individual PMT hits themselves
along with microsecond-level timestamps. To invoke list mode instead of PHA:

```python
base.set_acq_mode_list()
sleep(5) # If needed - only first time after changing HV
base.start()
```

Now don't dilly-dally before reading out the list buffer - it's only 128k elements 
(or maybe bytes in which case it's only 32k elements!) deep. The base can run 
sustained over 32 ksps so you have about a second to start draining the buffer:

```python
hits = []
while some_condition_is_true:
    while len(new_hits := base.hits) > 0: hits += new_hits
```

I've used the Python 3.8+ walrus operator to enter a tight loop that drains the
device's internal buffer. The device reads are limited to 4096 bytes so you need
to ensure that there are not hits left in the buffer, hence that inner read loop.

The hits themselves are 32-bit integers which encode time and charge. There are
actually two kinds of data that are encountered in the hit list readout:

* PMT hits have bit 31 clear. In this case bits 30-21 are 10-bit ADC / charge and
bits 20-0 are time in units of microseconds (according to the not very precise
local oscillator on the digiBase);
* Time rollover words have bit 31 set. In this case bits 30-0 are the time epoch
in microseconds. Because the PMT hits only have 21 bit these rollover markers 
allow for rollover correction.

Here's an example of how to use a hit list populated with PMT hits and rollover
words:

```python
t20 = 0
hit_times = []
hit_q = []
for h in hits:
    if h & 0x8000_0000:
        t20 = h & 0x7fff_ffff
    else:
        t = t20 + (h & 0x001f_ffff)
        hit_times.append(t)
        hit_q.append((h >> 21) & 0x3ff)
```

## Using the External Gate
Hit collection can be suppressed using a TTL-level signal connected to
the SMA input on the base. This suppression occurs for both PHA _and_
list mode acquisitions. The external gate mode determines how the base
responds to this external gate:

* When mode = `ExtGateMode.OFF`, the external gate is ignored, _i.e._ 
  acquisition continues.
* When mode = `ExtGateMode.ENABLED`, a TTL low on the external gate
  will suspend the data acquisition. As far as I can tell, this 
  also _stops_ the livetime counter as well as the microsecond
  timestamp counter in list acquisition mode.
* When mode = `ExtGateMode.COINCIDENCE`, TTL low will stop hit
  collection (i.e. no hits will fill the MCA channels in PHA
  mode and hits will not appear in list mode), however the livetime